Home

Specifications

Specifications
 
Particle size range 0.007 – 10 µm with filter stage
0.030 – 10 µm standard
Number of size classes 12
Sample flow rate 10 or 30 l/min
Impactor dimensions Ø 65 x 300 mm
ELPI dimensions H 570 x W 400 x D 230 mm
Collection plate diameter 25 mm
Unit weight 35 kg
Lowest stage pressure 100 mbar
Pump requirements* 7 m3/h at 100 mbar (10 lpm)21 m3/h at 100 mbar (30 lpm)
Operating temperature 5 – 40 ºC
Operating humidity 0- 90 % RH Non-condensing
Response time < 5 sec
Computer requirements Pentium III 500 MHz, 128 MB RAM, MS-Windows NTTM, 2000TM or XPTM
 

*Suitable pumps available at Dekati Ltd.

Each ELPI™ impactor is individually calibrated before delivery; the calibration includes detailed determination of the exact sample flow rate and D50% values. The nominal values are presented in the table below.

 
Stage
D50%
 
[µm]
13
10
12
6.8
11
4.4
10
2.5
9
1.6
8
1.0
7
0.65
6
0.40
5
0.26
4
0.17
3
0.108
2
0.06
1
0.030
Filter
0.007