| Particle size range |
0 – 1.2 µm |
| Number of Stages |
6 impactor stages + one mobility channel |
| Volumetric flow rate |
10 lpm (nominal) |
| Lowest stage pressure |
100 mbars absolute pressure |
| Operation temperature |
5-40ºC
|
| Operation humidity |
0-90 % R.H. |
| Response time |
< 2 seconds |
| Concentration range |
From 1 to 1000 (5000) µg/m3 (continuous / short time) |
| DMM unit dimensions |
Outer cabin:
W x L x H 560 x 420 x 300 mm
Installable to 19” rack, 6 U height:
W x H x L 450 x 266 x 400 mm |
| Unit weight |
40 kg
|
|
Inlet/Outlet
|
G 3/8” thread, 10 mm male pipe connector
G 3/8” thread, NW-16 –connector |
| Input/output |
RS-232 serial connector for computer interface
0-5 V analog mass concentration output
2 0-5 V analog input signals |
| Required accessories |
Vacuum pump (e.g. IA-904) and a suitable vacuum hose (IA-203)
Computer for instrument control
Sampling system |
| Computer requirements |
Pentium III processor, MS-WINDOWS 2000 SP4, XP or Vista |
| Pump requirements |
7 m³/h at 100 mbar |