Home

Specifications


Specifications
 
Nominal airflow
10 or 30 lpm
Particle size range
0.030-10µm standard with filter stage extended < 0.030 µm
Number of stages
13 standard 14 with filter stage (13-10)
Collection plate size
Ø 25 mm for impactor stages Ø 47 mm for the filter stage
Inlet
NW16
Outlet
NW16
Operation temperature
0-200 ºC
Weight
5 kg
Dimensions
Ø 65 x 300 mm
Material
Stainless steel
Pump requirements* 7 m3/h at 100 mbar (10 lpm) 21 m3/h at 100 mbar (30 lpm)
 
Each DLPI impactor is individually calibrated before delivery; the calibration includes detailed determination of the exact sample flow rate and D50% values. The nominal values are presented in the table below.
 
 
Stage
D50%
 
[µm]
13
10
12
6.8
11
4.4
10
2.5
9
1.6
8
1.0
7
0.65
6
0.40
5
0.26
4
0.17
3
0.108
2
0.06
1
0.030
Filter
~ 0